Ultrahigh Resolving Power Ion Mobility Spectrometry with a Simple Pulser Circuitry

Marc-Aurèle Boillat,Peter C. Hauser
DOI: https://doi.org/10.1021/acs.analchem.4c04881
IF: 7.4
2024-11-30
Analytical Chemistry
Abstract:The pulsing circuitry for high resolving power drift-tube ion-mobility spectrometry is based on three avalanche photodiodes. These are switched on by illumination through optical fibers, which provide electrical insulation of the driver circuitry from the high voltage. The setup was tested with a series of quaternary ammonium ions introduced with an electrospray ion source. Two instruments with drift tubes of 10 and 30 cm length were employed and a separation voltage of up to 23.7 kV. Resolving...
chemistry, analytical
What problem does this paper attempt to address?