Abstract:A dual ion mobility spectrometer is developed to quickly detect and analyze complex molecules (such as explosives, drugs, DNA compounds and enzymes), which simplifies the large, complex circuit required when using transformers to measure a weak signal in a high voltage. The spectrometer has an integrated structure, is simple to manufacture and is easily operated. Positive and negative drift tubes in the spectrometer can detect both positive and negative ion affinity molecules. The weak signal of 10-13 C on the Faraday plate is driven by an AC coupling circuit. The spectrometer is driven by a 12 V DC, TTL pulse and can be easily controlled. The spectrometer can work continuously with many complex molecules detected by the system. The sensitivity when used to detect TNT (a reference molecular used world wide) is better than 1 ng and the relative resolution for TNT can reach 21 when the drift tube is 50 mm long and 32 when the drift tube is 90 mm long.