Charge Detection Mass Spectrometry for Megadalton Polymer Characterization and Measurement of Electrospray-Generated Charged Droplet Dynamics with a New "Direct Visualization of the Rayleigh Limit" Approach to Aid in m / z Calibration

David Hrabovsky,Bérengère Argence,Denis Lesage,Philippe Colomby,Michel Surugue,Richard B. Cole
DOI: https://doi.org/10.1021/acs.analchem.4c00055
IF: 7.4
2024-04-24
Analytical Chemistry
Abstract:A charge detector has been constructed and mounted inside the vacuum housing of a commercial mass spectrometer (Micromass-Waters Quattro I, Waters Corp., Manchester, UK). The in-house built single-pass charge detector is composed of a designed, complete electronics system that includes a low-noise charge amplifier. Communication to the data acquisition system was enabled, and analog and digital filters were devised, followed by their tuning and programming. Data treatment scripts for data...
chemistry, analytical
What problem does this paper attempt to address?