Pop-up Langmuir probe diagnostic in the water cooled divertor of Wendelstein 7-X

A. Pandey,S. Bohm,A. Carls,C. Cordes,M. Endler,J. Fellinger,S. Freundt,K. Gallowski,K. Hammond,D. Hathiramani,G. Isberner,J. P. Kallmeyer,M. Krause,J. Kügler,M. Otte,T. S. Pedersen,D. Rondeshagen,J. Ruhnau,T. Schröder,T. Sieber,J. Wendorf,
DOI: https://doi.org/10.1063/5.0188738
IF: 1.6
2024-04-01
Review of Scientific Instruments
Abstract:The design, development, and successful implementation of pop-up Langmuir probes installed in the water-cooled divertor of W7-X are described. The probes are controlled by drive coils (actuators) installed behind the divertor plates. These drive coils make use of the magnetic field in W7-X to move the probe tips into and out of the plasma. The drive coils were installed in the vacuum vessel after extensively testing the durability of the coils and analyzing the criteria for safe operation. The probe design is carefully tailored for each of the 36 probe tips in order to be suitable for the different magnetic field configurations used in W7-X and ensure that the probes do not present leading edges to the magnetic flux tubes. An electronic bridge circuit is used for measurement to compensate for the effects of signal propagation time on the long cable lengths used. The diagnostic is integrated with the segment control of W7-X for automated operation and control of the diagnostic. The evaluation of the results from the plasma operation is presented after accounting for appropriate sheath expansion for negative bias voltage on the probes.
instruments & instrumentation,physics, applied
What problem does this paper attempt to address?
The paper aims to address the issue of Langmuir probe measurements on the High Heat Flux (HHF) divertor of the Wendelstein 7-X (W7-X) tokamak device. Specifically, the paper describes the following points: 1. **Design and Development**: The paper introduces the design, development, and successful implementation of a pop-up Langmuir probe in the water-cooled divertor. These probes are controlled by drive coils (actuators) installed behind the divertor plates, utilizing the magnetic field in W7-X to move the probes in and out of the plasma. 2. **Probe Material Selection**: Since the probes need to withstand extremely high heat loads (approximately 100-200 MW/m²), it is essential to choose materials that can endure high temperatures and have good electrical conductivity. Tungsten was ultimately selected as the probe material because it has a high melting point (3700 K) and good thermal conductivity. 3. **Probe Positioning**: The position of the probes was carefully designed to ensure accurate measurement of plasma parameters under different magnetic field configurations and to avoid contact between the probes and the flux tubes. 4. **Safety Assessment**: The paper provides a detailed analysis of the potential risks associated with using high-Z materials (such as tungsten), including the issue of radiation collapse caused by tungsten impurities entering the plasma core. Simulations and experimental data demonstrate that even in the event of probe damage, the impact of tungsten impurities is controllable. In summary, the main objective of the paper is to achieve accurate measurement of plasma parameters under high heat flux conditions in W7-X, while ensuring the reliability and safety of the system.