Design of a depth-encoding detector using 2 mm scintillator block and light guides with different reflector patterns for each layer: a simulation study

Seung-Jae Lee,Byungdu Jo,Cheol-Ha Baek
DOI: https://doi.org/10.1007/s40042-024-01035-5
2024-02-27
Journal of the Korean Physical Society
Abstract:A detector measuring the interaction depth of four layers was designed using a 2 mm-sized scintillation pixel and a light guide to which different reflector patterns were applied for each layer. DETECT2000 simulation was performed to evaluate the interaction depth layer measurement of the designed detector. The size of the scintillation pixel was 2 mm × 2 mm × 5 mm, and the light guide was composed of a total of four layers using a thickness of 2 mm. A gamma ray was interacted at every scintillation pixel to generate corresponding light, which was collected by the photosensor. The collected light signals were summed by giving weights according to the location of the photosensor pixel. The summed signal was reconstructed into a flood image through the Anger equation. As a result, all scintillation pixels appeared separated from each other in the flood image. This indicates that the interaction depth layer can be measured as four layers. When the designed detector is used in a PET system for small animals, a small scintillation pixel with a size of 2 mm is used and four layers of interaction depth can be distinguished, so it is considered that very good spatial resolution can be achieved throughout the field of view.
physics, multidisciplinary
What problem does this paper attempt to address?