Construction and evaluation of a large radiation detector for Positron Emission Tomography applications

R.R. Raylman,A.V. Stolin,G. Jaliparthi,P.F. Martone
DOI: https://doi.org/10.1088/1748-0221/19/04/p04034
2024-04-28
Journal of Instrumentation
Abstract:Large arrays of pixelated scintillator potentially have application in economical construction of PET scanners. In this investigation, we constructed and evaluated a detector with an active area of 32.26 × 13.47 cm 2 . It is based on a 218 × 91 array of 1.4 × 1.4 × 15 mm 3 LYSO elements (pitch 1.48 mm). Scintillation light is detected with a 5 × 12 array of silicon photomultipliers (SiPM) arrays. Each array consists of an 8 × 8 array of 3 × 3 mm 2 (pitch 3.35 mm) SiPMs. Performance of these devices are enhanced and stabilized by cooling them. Testing revealed that the detector was able to detect 90% of the theoretically detectable 511 keV photons. The resolvability index (a measure of the ability to identify individual detector elements from background) is 0.24 ± 0.04. Additionally, the average energy resolution for the complete detector is 18.3%. These results compare well with those reported for much smaller detector modules.