Evaluation of a novel robotic testing method for stability and kinematics of total knee arthroplasty

Sander R. Holthof,Mick Rock,Richard van Arkel,Angela Brivio,David Barrett,Andrew A. Amis
DOI: https://doi.org/10.1002/ksa.12516
2024-10-26
Knee Surgery Sports Traumatology Arthroscopy
Abstract:Purpose This work developed a novel preclinical test of total knee replacements (TKRs) in order to explain TKR instability linked to patient dissatisfaction. It was hypothesized that stability tests on the isolated moving prostheses would provide novel comparative data on the stability and kinematics among TKR designs. Methods Three TKR designs, DePuy Synthes Attune MS, Stryker Triathlon and Zimmer Biomet Persona MC, were assessed using a robotic arm while flexing–extending 0–140°. Tests imposed 710 N body weight combined with three tibial loads: no anterior–posterior (AP) force, 90 N anterior or 90 N posterior force. Other load effects were minimized and the kinematics was recorded. Each implant was tested six times to investigate the repeatability of the method. Data were analysed using statistical parametric mapping with one‐way analysis of variance (ANOVA). If significance was found (p
surgery,orthopedics,sport sciences
What problem does this paper attempt to address?