Abstract:Total knee arthroplasty aims to relieve pain and restore function in the affected joint through artificial implants. Despite advances in design and surgical techniques, there are complications associated with surgery, especially concerning the patella. The use of musculoskeletal models in orthopedic surgery allows for objective prediction of postoperative function and optimization of results for each patient. To ensure that simulations are trustworthy and can be used for predictive purposes, comparing simulation results with experimental data is crucial. Although progress has been made in obtaining 3D bone geometry and estimating contact forces, validation of these predictions has been limited due to the lack of direct in vivo measurements and the economic and ethical constraints associated with available alternatives. In this study, an existing commercial surgical training station was transformed into a sensorized test bench. The use of 3D-printed models and sensors allowed low-cost and reproducible experimental validation of computer simulation (patellar movement and forces) while avoiding ethical issues.