Visualization of radiotracers for SPECT imaging using a Timepix detector with a coded aperture

V. Rozhkov,G. Chelkov,I. Hernández,O. Ivanov,D. Kozhevnikov,A. Leyva,A. Perera,D. Rastorguev,P. Smolyanskiy,L. Torres,A. Zhemchugov
DOI: https://doi.org/10.1088/1748-0221/15/06/P06028
2020-06-26
Journal of Instrumentation
Abstract:The work shows the ability to visualize radiotracers used in SPECT with a system based on a coded aperture mask and a hybrid pixel Timepix detector with the CdTe sensor. Characterization of the system using X-rays and radioactive sources confirms that the spatial resolution of less than 1 mm with a field of view 3 cm × 3 cm can be achieved. The results of a simulation study to determine the expected spatial resolution of the system in the focal plane for the various radionuclides is presented. The possibility of using this system with a thin (1 mm) coded aperture mask for reconstructing images of gamma emitters with the energy up to 180 keV is demonstrated.
instruments & instrumentation
What problem does this paper attempt to address?