Electrochemical performance of a three-layer electrode based on Bi nanoparticles, multi-walled carbon nanotube composites for simultaneous Hg(II) and Cu(II) detection

Qiwen Bao,Gang Li,Zhengchun Yang,Peng Pan,Jun Liu,Jiayuan Chang,Jun Wei,Ling Lin
DOI: https://doi.org/10.1016/j.cclet.2020.06.021
IF: 9.1
2020-10-01
Chinese Chemical Letters
Abstract:<p>Electrochemical analysis is a promising technique for detecting biotoxic and non-biodegradable heavy metals. This article proposes a novel composite electrode based on a polyaniline (PANi) framework doped with bismuth oxide multi-walled carbon nanotubes (Bi ) for the simultaneous detection of multiple heavy metal ions. Composite electrodes are prepared on screen-printed electrodes (SPCEs) using an efficient dispensing technique. We used a SM200SX-3A dispenser to load a laboratory-specific ink with optimized viscosity and adhesion to draw a pattern on the work area. The SPCE was used as substrate to facilitate cost-effective and more convenient real-time detection technology. Electrochemical techniques, such as cyclic voltammetry and differential pulse voltammetry, were used to demonstrate the sensing capabilities of the proposed sensor. The sensitivity, limit of detection, and linear range of the PANi-Bi electrode are 2.57×10<sup>2</sup> µA L µmol<sup>-1</sup> cm<sup>-2</sup>, 0.01 nmol/L, and 0.01 nmol/L-5 mmol/L and 0.15×10<sup>-1</sup> µA L µmol<sup>-1</sup> cm<sup>-2</sup>, 0.5 nmol/L, and 0.5 nmol/L-5 mmol/L for mercury ion (Hg(II)) and copper ion (Cu(II)) detection, respectively. In addition, the electrode exhibits a good selectivity and repeatability for Hg(II) and Cu(II) sensing when tested in a complex heavy metal ion solution. The constructed electrode system exhibits a detection performance superior to similar methods and also increases the types of heavy metal ions that can be detected. Therefore, the proposed device can be used as an efficient sensor for the detection of multiple heavy metal ions in complex environments.</p>
chemistry, multidisciplinary
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to develop an efficient, sensitive and selective electrochemical sensor for the simultaneous detection of mercury ions (Hg(II)) and copper ions (Cu(II)) in water bodies. Specifically, the paper proposes a three - layer electrode structure based on bismuth nanoparticles (Bi NPs), multi - walled carbon nanotubes (MWCNTs) and polyaniline (PANi) composites, aiming to improve the detection performance of these two heavy metal ions. ### Main research background and problems 1. **Environmental and health impacts**: - Mercury ions (Hg(II)) and copper ions (Cu(II)) are extremely harmful to the environment and human health. Mercury ions are highly toxic and may cause central nervous system disorders, speech disorders, movement disorders and muscle atrophy. Excessive copper ions may cause kidney and liver function disorders. - Traditional heavy metal ion detection methods such as atomic absorption spectrometry (AAS), inductively coupled plasma - mass spectrometry (ICP - MS), ultraviolet - visible spectroscopy (UV - Vis) and surface - enhanced Raman spectroscopy (SERS) are accurate but costly and complex to operate, and are not suitable for on - site rapid detection. 2. **Advantages of electrochemical analysis**: - Electrochemical analysis methods have attracted wide attention in heavy metal ion detection because of their simplicity, low cost, low sample preparation requirements, rapid analysis, and ease of miniaturization and integration. 3. **Limitations of existing materials**: - A single material has limitations in detection performance, such as deficiencies in conductivity, specific surface area and stability. Therefore, new composite materials need to be developed to improve the detection performance. ### Main contributions of the paper - **Composite material design**: - A composite material based on polyaniline (PANi), bismuth nanoparticles (Bi NPs) and multi - walled carbon nanotubes (MWCNTs) is proposed, and a three - layer electrode structure is prepared by the layer - by - layer assembly technique. - This composite material has good conductivity, a large specific surface area and excellent electrocatalytic performance, and can significantly improve the detection sensitivity and selectivity of Hg(II) and Cu(II). - **Performance evaluation**: - The sensing performance of the composite electrode is evaluated in detail by electrochemical techniques such as cyclic voltammetry (CV) and differential pulse voltammetry (DPV). - The results show that the detection limits (LOD) of the composite electrode for Hg(II) and Cu(II) are 0.01 nmol/L and 0.5 nmol/L respectively, the linear ranges are 0.01 nmol/L to 5 mmol/L and 0.5 nmol/L to 5 mmol/L respectively, and the sensitivities are 2.57×10² µA L µmol⁻¹ cm⁻² and 1.48×10⁻² µA L µmol⁻¹ cm⁻² respectively. - **Potential for practical applications**: - The composite electrode shows good selectivity and stability in complex heavy metal ion solutions and has the potential to be applied to portable heavy metal ion detection devices. In summary, through the design and preparation of a new type of composite electrode material, this paper successfully solves the problems existing in traditional detection methods and provides an effective solution for environmental monitoring and water quality safety.