The Characterization of A Double-Side Nanoband Electrode Array and Its Application to Heavy Metals Detection
Hui-Xin Zhao,Wei Cai,Da Ha,Hong-Sun Guo,Ping Wang
DOI: https://doi.org/10.1166/sl.2011.1618
2011-01-01
Sensor Letters
Abstract:Contamination of aquatic environment by heavy metals is serious because of the toxicity of such metals on the whole ecological system. This paper describes a double-side nanoband electrode array, which consisted of gold and platinum electrodes on the opposite surface and was used for heavy metals detection. The fabrication quality of the array was validated by scanning electron microscope (SEM) and energy dispersive X-ray (EDX). Cyclic voltammetry was utilized to characterize the electrochemical behavior of the array. Well correlation was found between experimental data and theoretical models. The analytical performance of the array was investigated using anodic stripping voltammetry, and detection limits below mu g.L(-1) level were obtained for lead and copper ions. The Au nanoband electrode array without plating mercury was applied to the detection of heavy metals in sea water, where agreement was found with the standard analytical method (Atomic Absorption Spectrometry, AAS). This paper proposed a mercury-free double-side nanoband electrode array, offing the possibility for in situ, on line heavy metals detection.
What problem does this paper attempt to address?