Microelectrode array sensor chip for detection of heavy metals in water pollution

赵会欣,万浩,蔡巍,哈达,王平
DOI: https://doi.org/10.3785/j.issn.1008-973X.2013.06.008
2013-01-01
Abstract:Based on stripping voltammetry, a novel gold microelectrode array (MEA) sensor chip with mercury film plated was used for simultaneously detection of trace heavy metal ions Zn2+, Cd2+, Pb2+ and Cu2+ in water pollution. The electrochemical behavior and the actual surface area of the MEA were investigated by cyclic voltammetry in K3[Fe(CN)6] solution. Electrochemical impedance spectrum (EIS) was utilized to examine the mercury deposition extent on the surface of the MEA. Mercury film plated MEA was then applied to simultaneously detect four heavy metal elements Zn2+, Cd2+, Pb2+ and Cu2+ in sample solutions using differential pulse anodic stripping voltammetry, where good linearities of the standard working curves were obtained for Cd2+, Pb2+ and Cu2+ with the detection limits of 0.1, 0.5 and 0.3 μg/L, separately. But the linearity of the standard working curve was poor for Zn2+, which was mainly caused by the excursion of the testing baseline and the inaccurate addition.
What problem does this paper attempt to address?