Smart Voltammetric Procedure In An Automatic Trace Metal Monitoring System For Expanding The Measurement Range Of A Gold-Band Microelectrode Array
Wen Zhang,Hao Wan,Qiyong Sun,Da Ha,Ping Wang,Dmitry Kirsanov,Andrey Legin
DOI: https://doi.org/10.1088/0957-0233/24/4/045801
IF: 2.398
2013-01-01
Measurement Science and Technology
Abstract:A smart voltammetric procedure in an original automatic trace metal monitoring system is presented in this study. Its aim is to expand the measurement range of a gold-band microelectrode array (MEA) for monitoring heavy metal ions such as Zn2+, Pb2+ and Cu2+. A voltammetric controller to execute a smart voltammetric procedure and a homemade mercury-free gold-band MEA are introduced. Three particular features are included in the smart procedure, which are self-calibration, multi-standard solution fluidics control and automatic adjustment of deposition time. The measurement range of a gold-band MEA is investigated by a traditional voltammetric set-up. Experiments show good linearity from 10 to 50 mu g L-1 for Zn2+, Pb2+ and Cu2+; the MEA detection limit for Zn2+, Pb2+ and Cu2+ is about 0.4, 1 and 0.4 mu g L-1, respectively. Moreover, self-calibration, multi-standard solution fluidics control and automatic adjustment of deposition time are introduced in smart voltammetric measurement with the method of standard additions. The upper concentration limit can be increased to twice that of conventional measurements; the detection limit for Zn2+, Pb2+ and Cu2+ can reach 0.1, 0.2 and 0.1 mu g L-1, respectively. In addition, this design has been validated by monitoring natural concentrations of Zn2+, Pb2+ and Cu2+ (West Lake, China) and comparing with atomic absorption spectrometry.