Wireless Instrumentation Design for Heavy Metals Detection in Water Based on Light Addressable Potentiometric Sensor

Xu WANG,Da HA,Wen ZHANG,Hao WAN,Qiyong SUN,Haibo LI,Fei SUN,Ping WANG
DOI: https://doi.org/10.3969/j.issn.1004-1699.2014.01.002
2014-01-01
Abstract:A novel system based on light addressable potentiometric sensor( LAPS) was proposed and fabricated for de-termination of heavy metals in water environment. Compared with other detection methods such as atomic fluorescence spectrophotometry,this type of instrument is much easier to use,with the characteristics of automatic range controlling and uncomplicated pretreatment process, etc. The instrument can communicate with computer via ARM wirelessly. Equipped with the independent fluidic system,it can detect pH and Pb2+ concentration in water samples with a fast and automatic response, which has been verified. The detection range of pH unit was 5 to 9, with a sensitivity of 37. 321 mV/pH and mean error of 4. 4%,which meets the requirement of pH detection. The detection range of Pb2+unit was 10-7 to 10-1 mol/L, with a sensitivity of 26. 43 mV/pPb2+ and average recovery rate of 86. 7%. In the meantime,the interference of Zn2+,Cu2+ and pH to Pb2+ detection has also been investigated.
What problem does this paper attempt to address?