Design of an Optical Instrument for Copper Ion Detection in Water Environment

Zhiwen LI,Haibo LI,Hao WAN,Qiyong SUN,Fei SUN,Jiawei TU,Ping WANG
DOI: https://doi.org/10.3969/j.issn.1004-1699.2016.01.025
2016-01-01
Abstract:An automated instrument was designed for heavy metal determination due to severe heavy metal contami-nation in water environment. The instrument can be used to specifically detect copper ion in water based on spectro-photometry. The instrument has the advantages of fast detection,simple operation and good repeatability,etc. Be-sides,specific fluid module was designed to implement auto-sampling during the detection. The waterway closed-loop detection module was also designed to achieve precise injection and error warning. In samples with different concentration,the instrument can adjust the laser intensity automatically based on the response of different sam-ples. The experiment result shows that the detection limit of the instrument is 0.006 mg/L,with good linearity in the range from 0.1 mg/L to 2.00 mg/L. Extraordinary performance was also obtained with both good accuracy less than 6.4%and precision less than 4.52%.