Polymer Waveguide Sensor Based on Evanescent Bragg Grating for Lab-on-a-Chip Applications

Zhenyu Zhang,Ahmad Abdalwareth,Günter Flachenecker,Martin Angelmahr,Wolfgang Schade
DOI: https://doi.org/10.3390/s24041234
IF: 3.9
2024-02-15
Sensors
Abstract:In this work, an evanescent Bragg grating sensor inscribed in a few-mode planar polymer waveguide was integrated into microchannel structures and characterized by various chemical applications. The planar waveguide and the microchannels consisted of epoxide-based polymers. The Bragg grating structure was postprocessed by using point-by-point direct inscription technology. By monitoring the central wavelength shift of the reflected Bragg signal, the sensor showed a temperature sensitivity of −47.75 pm/K. Moreover, the functionality of the evanescent field-based measurements is demonstrated with two application examples: the refractive index sensing of different aqueous solutions and gas-phase hydrogen concentration detection. For the latter application, the sensor was additionally coated with a functional layer based on palladium nanoparticles. During the refractive index sensing measurement, the sensor achieved a sensitivity of 6.5 nm/RIU from air to 99.9% pure isopropyl alcohol. For the gas-phase hydrogen detection, the coated sensor achieved a reproducible concentration detection up to 4 vol% hydrogen. According to the reported experimental results, the integrated Bragg-grating-based waveguide sensor demonstrates high potential for applications based on the lab-on-a-chip concept.
engineering, electrical & electronic,chemistry, analytical,instruments & instrumentation
What problem does this paper attempt to address?
The paper aims to address the following issues: 1. **Temperature Sensitivity Measurement**: A polymer waveguide-based Bragg grating sensor (RWBG) was studied, and its temperature sensitivity was characterized under different environmental temperatures. Experimental results show that the Bragg center wavelength sensitivity of the sensor to temperature changes is −47.75 pm/K. 2. **Refractive Index Detection**: The functionality of the sensor for refractive index measurement in liquid phase was demonstrated and verified using isopropanol-water solutions of different concentrations. The sensor achieved a refractive index detection sensitivity of 6.5 nm/RIU from air to 99.9% pure isopropanol solution. 3. **Hydrogen Concentration Detection**: By coating palladium nanoparticles on the sensor surface, the detection of hydrogen concentration in the gas phase was achieved. This coated sensor can detect hydrogen concentrations up to 4 volume percent (vol%) with good repeatability. In summary, this study demonstrates a Bragg grating sensor integrated on a microfluidic chip with high sensitivity for temperature, refractive index, and hydrogen detection, showing great potential for lab-on-a-chip applications.