A study on the validation of a single-particle measurement system for leakage gamma-ray decay curves

Hao Xi,Jie Wen,Meng Li,Hui Gao,Haojun Zhou,Qilin Xie,Taiping Peng
DOI: https://doi.org/10.1016/j.nimb.2024.165262
2024-02-01
Abstract:Using gamma rays as a probe has garnered increasing attention in dynamic measurement as the potential of obtaining an earlier fitting starting point and less diffusion time. We developed a single-particle measurement system based on the randomly pulsed source method. And we presented a comprehensive validation process by measuring the leakage ray decay curve of borated polyethylene samples. The system primarily comprised an organic liquid scintillator detector and a waveform digitizer. We carefully selected source intensity and detector efficiency to meet the single-particle condition. Additionally, we implemented energy and time calibration, pulse shape discrimination, and time-dependent coincidence to enhance signal quality by reducing noise. The experimental gamma-ray decay curves closely match the Monte Carlo simulation results, demonstrating the system's feasibility and reliability. The complete verification process provides convenience for more accurate measurement with subcritical assemblies in future research.
physics, nuclear, atomic, molecular & chemical,nuclear science & technology,instruments & instrumentation
What problem does this paper attempt to address?