Efficiency Measurement of NaI(Tl) Detector for a Long Tube Volume Source of 6.13mev Gamma-Ray

ZR Shi,K Xu,SY Ding,SQ Wang,ZR Guo,LZ Zuo
DOI: https://doi.org/10.3321/j.issn:0254-3052.2005.02.005
2005-01-01
Abstract:The relation of the N-16 leak rate of the nuclear reactor to the efficiency and counting rate of gamma-ray detector, velocity of high pressure vapor, is deduced. A conclusion of the efficiency independent on the velocity of high pressure vapor and decay constant of N-16 is drawn. Experimentally the space distribution of the relative efficiencies and the absolute efficiency at a reference point are measured with PuC source emiting 6.13 MeV gamma ray detected by NaI(Tl) counter. The measured efficiency is consistent with Monte Carlo calculation in 7%. The Compton small angle scattering contribution to the convergence length and the effect of the energy difference between 7.12 MeV and 6.13 MeV gamma ray on measured N-16 leak rate are discussed.