Reducing current noise in cryogenic experiments by vacuum-insulated cables

E. Mykkänen,J. S. Lehtinen,A. Kemppinen,C. Krause,D. Drung,J. Nissilä,A. J. Manninen
DOI: https://doi.org/10.1063/1.4963904
IF: 1.6
2016-10-01
Review of Scientific Instruments
Abstract:We measure the current noise of several cryogenic cables in a pulse tube based dilution refrigerator at frequencies between about 1 mHz and 50 kHz. We show that vibration-induced noise can be efficiently suppressed by using vacuum-insulated cables between room temperature and the 2nd pulse tube stage. A noise peak below 4 fA at the 1.4 Hz operation frequency of the pulse tube and a white noise density of 0.44 fA/Hz in the millihertz range are obtained.
instruments & instrumentation,physics, applied
What problem does this paper attempt to address?