Magnetic cooling for microkelvin nanoelectronics on a cryofree platform

M. Palma,D. Maradan,L. Casparis,T.-M. Liu,F. N. M. Froning,D. M. Zumbühl
DOI: https://doi.org/10.1063/1.4979929
IF: 1.6
2017-04-01
Review of Scientific Instruments
Abstract:We present a parallel network of 16 demagnetization refrigerators mounted on a cryofree dilution refrigerator aimed to cool nanoelectronic devices to sub-millikelvin temperatures. To measure the refrigerator temperature, the thermal motion of electrons in a Ag wire-thermalized by a spot-weld to one of the Cu nuclear refrigerators-is inductively picked-up by a superconducting gradiometer and amplified by a SQUID mounted at 4 K. The noise thermometer as well as other thermometers are used to characterize the performance of the system, finding magnetic field independent heat-leaks of a few nW/mol, cold times of several days below 1 mK, and a lowest temperature of 150 μK of one of the nuclear stages in a final field of 80 mT, close to the intrinsic SQUID noise of about 100 μK. A simple thermal model of the system capturing the nuclear refrigerator, heat leaks, and thermal and Korringa links describes the main features very well, including rather high refrigerator efficiencies typically above 80%.
instruments & instrumentation,physics, applied
What problem does this paper attempt to address?