Simultaneous multinuclear MRI via a single RF channel

Mohammad Rasool Vaezi K.,Jan G. Korvink,Mazin Jouda
DOI: https://doi.org/10.1016/j.jmr.2024.107782
IF: 2.734
2024-10-12
Journal of Magnetic Resonance
Abstract:Magnetic resonance imaging (MRI) stands as one of the most powerful noninvasive and non-destructive imaging techniques, finding extensive utility in medical and industrial applications. Its ability to acquire signals from multiple nuclei grants it additional levels of strength by providing multi-dimensional datasets of the object under test. However, this typically requires dedicated hardware to detect each nucleus. In this paper, we report on the use of a digital lock-in amplifier to perform simultaneous multi-nuclear MRI with a single physical radio frequency (RF) channel. While we showcase this concept by demonstrating the results of fully parallel (TX and RX) 1 Hand 19 FMRI images, we emphasize that it is not limited to two nuclei but can accommodate more nuclei with no extra cost on the hardware or scan time. The scalability is virtually unlimited, constrained only by the processing speed of the digital unit. Furthermore, we demonstrate that the quality of parallel imaging with SNR of 54 is comparable to the commercial single channel with SNR of 43. Thus with no reduction in imaging quality, the proposed concept promises a tremendous reduction in scan time, system complexity, and hardware costs.
spectroscopy,physics, atomic, molecular & chemical,biochemical research methods
What problem does this paper attempt to address?