Chemical characterization of ultra high molecular weight polyethylene based tibial inserts after ethylene oxide sterilization

Ogün BOZKAYA
DOI: https://doi.org/10.34088/kojose.1179821
2022-12-20
Kocaeli Journal of Science and Engineering
Abstract:Ultra high molecular weight polyethylene (UHMWPE) has become the gold standard for total joint replacements such as tibial inserts because of its chemical inertness, superior mechanical properties, and biocompatibility. Ethylene oxide sterilization is one of the most common and effective methods used, especially for sterilization of polyethylene based polymeric implants. However, variable sterilization conditions can cause a change in the chemical structure of the polymeric material, which affects its mechanical properties and lifetime. The aim of this study is to investigate whether the chemical structure of UHMWPE tibial inserts sterilized with ethylene oxide undergoing certain conditions remains the same. Chemical characterization studies were performed with Fourier transform infrared spectroscopy, Raman spectroscopy, X-ray diffractometer, differential scanning calorimetry, thermogravimetric analysis, mass spectrometry and elemental analysis techniques recommended for polymeric materials in ISO 10993-8:2020 standard. Characterization results revealed that there was no significant change in molecular structure, crystallinity, elemental composition of UHMWPE materials after ethylene oxide sterilization. These results can provide assurance that tibial inserts keep their physical, chemical, and mechanical properties after sterilization.
English Else
What problem does this paper attempt to address?