An LC-MS/MS method for quantitation of methylparaben, benzophenone, and pharmaceutical compounds from domestic wastewater

Juliana Cristina Barreiro,Anna Patrícya Florentino,Izadora Liranço Furlani,Gustavo H. R. Silva,Quezia Bezerra Cass
DOI: https://doi.org/10.3389/frans.2022.941883
2022-09-12
Frontiers in Analytical Science
Abstract:An analytical method was developed to quantify a mixture of acetaminophen, metoprolol, methylparaben, carbamazepine, naproxen, estrone, estradiol, diclofenac, benzophenone, ibuprofen, progesterone, and mefenamic acid from domestic wastewater samples. To match fast and efficient chromatographic separation for different classes of compounds, an automated scouting liquid chromatographic system was associated with the experimental design produced by the DryLab® software. HLB cartridges were used to extract the analytes from the sample matrix, which was followed by detection and quantitation by liquid chromatography-tandem mass spectrometry (LC-MS/MS). The lowest detection limits were found for acetaminophen and carbamazepine (0.625 pg L −1 ) and metoprolol (0.4 pg L −1 ).
What problem does this paper attempt to address?