Simultaneous determination of parabens and steroid estrogens in sewage water using gas chromatography-mass spectrometry

Liang WANG,Qianhui MAO,Shoujun YUAN,Deying HUANG,Jibiao ZHANG
DOI: https://doi.org/10.7524/j.issn.0254-6108.2016.01.2015081901
IF: 4.236
2016-01-01
Environmental Chemistry
Abstract:A method was developed for the simultaneous detamination of seven parabens and five steroid estrogens in sewage water using gas chromatography-mass spectrometry coupled by solid phase extraction. The sample derivatization was optimized and the best derivatization was obtained when target compounds reacted with bis ( trimethylsilyl ) trifluoroacetamide at 40 ℃ for 60 min. The optimum extraction conditions were as follow: HC-C18 cartridge was selected and the pH value of 7 was optimized for samples extraction, then ethyl acetate:acetone (1:1, V:V) was used for elution. Under the optimized conditions, the linear detection range of 12 analytes was 20—2000 ng·L-1( r>0.990). The limit of detection was 1.8—5.7 ng·L-1, and the recovery of 82.2-128.5%was obtained for most of the analytes, and the relative standard deviation was 4.4%—21.2%.
What problem does this paper attempt to address?