Numerical simulation and biomechanical analysis of locking screw caps on clavicle locking plates

Dae-Geun Kim,Soo Min Kim,Yoonkap Kim
DOI: https://doi.org/10.1097/md.0000000000029319
IF: 1.6
2022-08-12
Medicine
Abstract:The risk of displaced and comminuted midshaft clavicle fractures is increased in high-energy traumas such as sport injuries and traffic accidents. Open reduction and plate fixation have been widely used for midshaft clavicle fractures. Among various plates for clavicle shaft fractures, superior locking compression plates (LCPs) have been mostly used. In plate fixation, nonunion caused by implant failure is the most difficult complication. The most common reasons for metal plate failure are excessive stress and stress concentration caused by cantilever bending. These causes were easily addressed using a locking screw cap (LSC).
medicine, general & internal
What problem does this paper attempt to address?