An Implantable Wireless Neural Interface System for Simultaneous Recording and Stimulation of Peripheral Nerve with a Single Cuff Electrode

Ahnsei Shon,Jun-Uk Chu,Jiuk Jung,Hyungmin Kim,Inchan Youn
DOI: https://doi.org/10.3390/s18010001
IF: 3.9
2017-12-21
Sensors
Abstract:Recently, implantable devices have become widely used in neural prostheses because they eliminate endemic drawbacks of conventional percutaneous neural interface systems. However, there are still several issues to be considered: low-efficiency wireless power transmission; wireless data communication over restricted operating distance with high power consumption; and limited functionality, working either as a neural signal recorder or as a stimulator. To overcome these issues, we suggest a novel implantable wireless neural interface system for simultaneous neural signal recording and stimulation using a single cuff electrode. By using widely available commercial off-the-shelf (COTS) components, an easily reconfigurable implantable wireless neural interface system was implemented into one compact module. The implantable device includes a wireless power consortium (WPC)-compliant power transmission circuit, a medical implant communication service (MICS)-band-based radio link and a cuff-electrode path controller for simultaneous neural signal recording and stimulation. During in vivo experiments with rabbit models, the implantable device successfully recorded and stimulated the tibial and peroneal nerves while communicating with the external device. The proposed system can be modified for various implantable medical devices, especially such as closed-loop control based implantable neural prostheses requiring neural signal recording and stimulation at the same time.
engineering, electrical & electronic,chemistry, analytical,instruments & instrumentation
What problem does this paper attempt to address?
The problems that this paper attempts to solve mainly focus on several key issues in the existing implantable wireless neural interface systems: 1. **Inefficient wireless power transfer and limited power supply**: Most implantable wireless devices are powered by inductive links or non - rechargeable batteries. The former (batteryless inductive link) requires a very short distance, which limits the wearer's range of motion and provides inefficient wireless power transfer; the latter (non - rechargeable battery) has a service life and requires surgical replacement. 2. **High power consumption problem in long - distance data communication**: Current data communication methods (such as induction, optical and radio links) have the problem that patients need to maintain a fixed posture to ensure correct data transmission, and are easily interfered by other systems in specific frequency bands. In addition, in the Industrial, Scientific and Medical (ISM) frequency band, electromagnetic energy is absorbed by human tissues, resulting in serious signal attenuation. 3. **Single - function problem of electrodes**: When it is necessary to perform nerve stimulation and recording simultaneously, experimenters usually need to use two independent devices (one as a stimulator and one as an amplifier), as well as two different types of electrodes (one for recording and the other for stimulation). This not only increases the complexity of device insertion into the body, but also the artifacts generated by high - current stimulation are likely to contaminate weak neural signals. To solve the above problems, this paper proposes a new implantable wireless neural interface system, which has the following characteristics: - **Efficient wireless power transfer circuit**: It complies with the Wireless Power Consortium (WPC) standard and can charge rechargeable batteries through wireless power transfer under the skin, improving power transfer efficiency and reducing charging time. - **Wireless data communication based on MICS band**: A radio - frequency link in the Medical Implant Communication Service (MICS) band (402 - 405 MHz) is adopted to achieve a longer wireless data communication distance and lower power consumption, avoiding the problem of strong absorption of signals by human tissues in the ISM band. - **Multi - function path controller for a single cuff electrode**: A new type of switching circuit is designed to allow a single cuff electrode to be used for nerve stimulation and recording simultaneously, reducing the number of required cables, cuff electrodes and electronic devices, and reducing the influence of stimulation artifacts on the recorded signal. Through these improvements, the system can achieve synchronous recording and stimulation of neural signals without increasing the in - vivo space occupation, thus making it possible for applications such as implantable neural prostheses for closed - loop control.