An Ultra-Low-Noise, Low Power and Miniaturized Dual-Channel Wireless Neural Recording Microsystem

Haochuan Wang,Qian Ma,Keming Chen,Hanqing Zhang,Yinyan Yang,Nenggan Zheng,Hui Hong
DOI: https://doi.org/10.3390/bios12080613
2022-01-01
Abstract:As the basic tools for neuroscience research, invasive neural recording devices can obtain high-resolution neuronal activity signals through electrodes connected to the subject’s brain. Existing wireless neural recording devices are large in size or need external large-scale equipment for wireless power supply, which limits their application. Here, we developed an ultra-low-noise, low power and miniaturized dual-channel wireless neural recording microsystem. With the full-differential front-end structure of the dual operational amplifiers (op-amps), the noise level and power consumption are notably reduced. The hierarchical microassembly technology, which integrates wafer-level packaged op-amps and the miniaturized Bluetooth module, dramatically reduces the size of the wireless neural recording microsystem. The microsystem shows a less than 100 nV/Hz ultra-low noise level, about 10 mW low power consumption, and 9 × 7 × 5 mm3 small size. The neural recording ability was then demonstrated in saline and a chronic rat model. Because of its miniaturization, it can be applied to freely behaving small animals, such as rats. Its features of ultra-low noise and high bandwidth are conducive to low-amplitude neural signal recording, which may help advance neuroscientific discovery.
What problem does this paper attempt to address?