Severe tricuspid regurgitation due to interactions with right ventricular permanent pacemaker or defibrillator leads

Cory R. Trankle,Zachary M. Gertz,Jayanthi N. Koneru,Vigneshwar Kasirajan,Patricia Nicolato,Hem L. Bhardwaj,Kenneth A. Ellenbogen,Gautham Kalahasty
DOI: https://doi.org/10.1111/pace.13369
2018-06-04
Pacing and Clinical Electrophysiology
Abstract:Although thought to be a rare event, permanent pacemakers and implantable cardioverter-defibrillators with right ventricular intracardiac leads have the potential to induce tricuspid valve dysfunction. Adverse lead-valve interactions can take place through a variety of mechanisms including damage at the time of implantation, leaflet pinning, or long-term fibrosis encapsulating the leaflet tissue. Clinical manifestations can display a wide range of severity, as well as a highly variable time span between implantation and hemodynamic deterioration. This review aims to describe the potential pathophysiologic effects of intracardiac device leads on the tricuspid valve, with a focus on ideal diagnostic strategies and treatment options once lead-induced valvular dysfunction is suspected.