New evaluation beamline for soft x-ray optical elements

Masato Koike,Kazuo Sano,Osamu Yoda,Yoshihisa Harada,Masahiko Ishino,Naoji Moriya,Hiroyuki Sasai,Hisataka Takenaka,Eric Gullikson,Stanley Mrowka,Masafumi Jinno,Yoshihiro Ueno,James H. Underwood,Takeshi Namioka
DOI: https://doi.org/10.1063/1.1445489
IF: 1.6
2002-03-01
Review of Scientific Instruments
Abstract:An evaluation system capable of measuring the wavelength and angular characteristics of the absolute reflectivity (or diffraction efficiency) of soft x-ray optical elements has been designed and constructed. The system was installed on a beamline (BL-11) of the AURORA, a superconducting compact storage ring, at the Synchrotron Radiation Center, Ritsumeikan University. To cover a wavelength range of 0.5 nm<λ<25 nm, this system incorporates two types of Monk–Gillieson monochromators. One is a conventional type equipped with three varied-line-spacing gratings, allowing a choice of two included angles. The other is a new type that employs a scanning mechanism based on surface normal rotation. The outline of the system and some preliminary experimental data obtained in the course of test runs are described.
instruments & instrumentation,physics, applied
What problem does this paper attempt to address?