Microvolume field-effect pH sensor for the scanning probe microscope

S. R. Manalis,E. B. Cooper,P. F. Indermuhle,P. Kernen,P. Wagner,D. G. Hafeman,S. C. Minne,C. F. Quate
DOI: https://doi.org/10.1063/1.125942
IF: 4
2000-02-21
Applied Physics Letters
Abstract:A pH sensitive scanning probe is realized by integrating a micron-sized field-effect sensor onto a cantilever designed for an atomic force microscope. The hybrid device, called a scanning probe potentiometer (SPP), is capable of measuring pH gradients over a sample surface. The device was used to profile the pH across a reservoir of laminar streams created by fluid flow in an array of microfluidic channels of varying pH. When a single SPP scanned, a 1.5 mm reservoir in a 10-channel array, the pH profile was measured in less than 1 min with a spatial resolution of 10 μm and sensitivity of less than 0.01 pH units.
physics, applied
What problem does this paper attempt to address?