Inductively Coupled Facility Qualification for Electron Transpiration Cooling Investigations

Andrew J. Morin,Rebecca Osborn,Jeff C. Schindler,Precious Jagun,Douglas G. Fletcher,Jason M. Meyers
DOI: https://doi.org/10.2514/6.2020-0921
2020-01-05
Abstract:In this work we present a strategy to measure temperature and electron population in an inductively coupled plasma facility used for hypersonic aerothermodynamic studies. Normally in our lab, temperature is inferred from two-photon LIF measurements of either N atom or O atom when nitrogen or oxygen is one of the major gases within the plasma. However, re- cent investigations of thermionic emission in high temperature plasmas required flows of nearly pure argon to be studied. As such a strategy to extract temperature from optical emission spectroscopy was explored here. This temperature data was then used to extract electron popu- lations from Stark broadening of hydrogen lines in the Balmer series. The current measurements appreciably overestimate electron number density when compared to equilibrium. While the work presented here shows promise there are more near-term improvements to be made before the technique is employed as standard lab measurement strategy.
What problem does this paper attempt to address?