Rugged Film Resistor Thermometer for the Measurement of Surface Temperatures

R. E. Thun,G. F. Caudle,E. R. Pasciutti
DOI: https://doi.org/10.1063/1.1717003
IF: 1.6
1960-04-01
Review of Scientific Instruments
Abstract:A resistor thermometer element is described consisting of a nickel or palladium film protected by an SiO or didymium fluoride coating. The films are vacuum deposited on a glass substrate with imbedded Kovar contacts. The element is capable of withstanding accelerations up to 100 000 G. It is shown that nickel or palladium films can be obtained with a temperature coefficient of resistance (TCR) considerably higher than previously reported. When corrected for the anomalous skin effect, a resistivity and a TCR approaching the bulk values were measured on films deposited at a rate of about 200 A/sec and annealed for 20 min at 450°C.
instruments & instrumentation,physics, applied