Design of a Capillary Viscometer System for the Determination of Polymer Solution Viscosity

Sadao Mori,Tatsuya Housaki
DOI: https://doi.org/10.1080/10236669608033354
1996-08-01
International Journal of Polymer Analysis and Characterization
Abstract:The viscometer consists of two capillaries and differential pressure transducer assemblies, an HPLC pump, a dampner, a six-port loop valve, and a differential pressure amplifier. Two capillaries are connected in series, solvent is fed with the pump through the dampner to the capillaries continuously, and a polymer solution is fed between the two capillaries. The increase of the differential pressure drop between the two capillaries, one for solvent and the other for polymer solution, is amplified and recorded. The solvent was tetrahydrofuran and the flow rate was 2 mL/min. Intrinsic viscosities of several polystyrenes were measured and the values were in good agreement with those in the literature. Intrinsic viscosity of a polystyrene measured by using a Ubbelohde viscometer agreed with the value obtained with the capillary viscometer. The coefficients of Mark-Houwink equation for polystyrene were K = 8.45 × 10−5 and a = 0.741 using polystyrene standards of narrow molecular weight distributions.