The Study of a New Method to Determine Copper Ion by Square-Wave Voltammetry-Extraction Iodometry at the Liquid/Liquid Interfaces

Changyan Shi,Shaoai Xie,Jinping Jia
DOI: https://doi.org/10.1155/2008/453429
2008-01-01
Journal of Automated Methods and Management in Chemistry
Abstract:A new method of indirect determination of Cu 2 + was developed based on square-wave voltammetry by the oxidation of iodide in organic solvent at the liquid/liquid (L/L) interface. The limit of detection for the determination of Cu 2 + in this method was found to be 5 × 10 −4 mol/L, and the concentration ranged up to 1 × 10 −2 mol/L gave a linear limiting current versus concentration response. For the same simulated wastewater, this method showed high accuracy compared with the result tested by sodium diethyldithiocarbamate extraction spectrophotometry. This approach could be applied to the indirect determination of the oxidative heavy metals in the industrial wastewater.
What problem does this paper attempt to address?