A combined freeze chamber and low temperature stage for an electron microscope

U. Valdrè,R. W. Horne
DOI: https://doi.org/10.1111/j.1365-2818.1975.tb03987.x
IF: 1.9522
1975-04-01
Journal of Microscopy
Abstract:A low‐temperature freeze‐drying or freezing preparation chamber and specimen stage has been designed and constructed for use with an electron microscope. The system allows biological specimens to be prepared under conditions of temperature and pressure and their subsequent direct transfer to a cold stage maintained at — 175°C within the microscope. Attachment of both the chamber and stage is a simple procedure and does not interfere with the normal electrical and mechanical operation of the microscope. The stage is cooled with the aid of a liquid nitrogen reservoir system and can be maintained at — 175°C for about 80 min without the use of any continuous cooling system. Test specimens have shown that a resolution of about 1.0–1.5 nm can be maintained at — 175°C for 80 min once thermal equilibrium has been established.
microscopy
What problem does this paper attempt to address?