Development of a Liquid-Helium Free Cryogenic Sample Holder with mK Temperature Control for Autonomous Electron Microscopy

X. He,R. Kostin,E. Knight,M.G. Han,J. Mun,I. Bozovic,C. Jing,Y. Zhu
DOI: https://doi.org/10.1016/j.ultramic.2024.114037
IF: 2.994
2024-09-10
Ultramicroscopy
Abstract:The automated and autonomous cryogenic transmission electron microscopy (Cryo-EM) demands a sample holder capable of maintaining temperatures below 10K with precise control, long holding times, and minimal helium use. Rising to this challenge, we initiated an ambitious project to develop a novel closed-cycle cryocooler-based cryogenic sample holder that operates without the use of liquid helium and the consumption of gaseous helium. This article details the design, construction, and experimental testing of the initial prototype, which achieves an ultimate temperature of 5.6K with exceptional stability close to 1mK, marking a clear advancement in cryo-EM holder development. While the prototype was not designed for atomic resolution imaging and thus lacks a sturdy support system to mitigate mechanical vibrations from the cryocooler's pulsed tube, this innovative approach successfully demonstrates proof of concept. It offers unprecedented capabilities for state-of-the-art cryogenic microscopy and microanalysis in materials and biological sciences.