Ultra-Cold Cryogenic TEM with Liquid Helium and High Stability

Emily Rennich,Suk Hyun Sung,Nishkarsh Agarwal,Maya Gates,Robert Kerns,Robert Hovden,Ismail El Baggari
2024-02-01
Abstract:Cryogenic transmission electron microscopy has revolutionized structural biology and materials science, but achieving temperatures below the boiling point of liquid nitrogen remains a long-standing aspiration. We introduce an ultra-cold liquid helium transmission electron microscope specimen holder, featuring continuous cryogen flow and vibration decoupling. This instrument is compatible with modern aberration-corrected microscopes and achieves sub-25 K base temperature, ${\pm}$2 mK thermal stability over many hours, and atomic resolution--setting the stage for a new era of cryogenic electron microscopy.
Applied Physics
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to achieve transmission electron microscope (TEM) imaging in an ultra - low - temperature environment below the boiling point of liquid nitrogen (77 K). Specifically, the authors developed an ultra - low - temperature cryo - transmission electron microscope sample holder cooled by liquid helium to overcome the challenges existing in current technologies under low - temperature conditions, such as large temperature fluctuations, mechanical vibrations, and short operation time. ### Main Problems and Solutions 1. **Achieving Lower Temperatures**: - Current technologies can usually only reach the temperature of liquid nitrogen (about 77 K) and cannot be further reduced to lower temperatures. The ultra - low - temperature environment is crucial for improving radiation tolerance and resolution, especially in single - particle cryo - electron microscopy (cryo - EM) and cryo - electron tomography. - By introducing a liquid - helium cooling system, the authors achieved a substrate temperature below 25 K and were able to maintain a stable temperature of 23 K within 4 hours. 2. **Improving Temperature Stability**: - Existing low - temperature TEM technologies have large temperature fluctuations, which affect the imaging quality and the reliability of experimental results. - The newly - designed sample holder achieved a temperature stability of ±2 mK, which is significantly better than existing instruments (estimated to be improved by 10 times). This high stability reduces sample drift and makes long - term experiments possible. 3. **Reducing Mechanical Vibrations**: - The rapid evaporation of liquid helium can cause mechanical vibrations and thermal instability, affecting the imaging quality. - By using flexible ultra - high - vacuum edge - welded bellows and rubber damping blocks, the new design effectively isolates the vibration of the helium transmission line and the room - temperature environment, ensuring the stability of the sample. 4. **Extending the Operation Time**: - Existing low - temperature sample holders have a short operation time due to small - volume Dewar flasks and require frequent coolant replenishment. - The new design, by means of continuous - flow liquid helium, makes the operation time only limited by the helium supply, greatly extending the experimental time. Consuming less than 2 liters of helium per hour makes long - term experiments possible. ### Application Prospects This new type of ultra - low - temperature cryo - transmission electron microscope sample holder provides new opportunities for studying biological samples and quantum materials. For example, it can be used to observe structural phase transitions that occur at low temperatures, such as the charge - density - wave (CDW) phenomenon in 2H - NbSe₂. In addition, it can also be applied in fields such as single - particle reconstruction, chemical spectroscopy, tomography, and ptychography, promoting the research progress of sensitive materials and quantum materials. ### Formulas and Figures The key parameters mentioned in the article include: - Substrate temperature: $< 25 \, \text{K}$ - Temperature stability: $\pm 2 \, \text{mK}$ - Operation time: more than 10 hours - Helium consumption rate: < 2 L/h These improvements enable the ultra - low - temperature cryo - transmission electron microscope technology to be more widely applied in biological and materials science research, especially in exploring the unique physical and chemical properties at low temperatures.