A compact ultrahigh vacuum scanning tunneling microscope with dilution refrigeration

Timofey Balashov,Michael Meyer,Wulf Wulfhekel
DOI: https://doi.org/10.1063/1.5043636
2018-06-12
Abstract:We have designed and built a scanning tunneling microscope (STM) setup for operation at millikelvin temperatures in ultra high vacuum. A compact cryostat with an integrated dilution refrigerator has been built, that allows measurements at a base temperature of 25 mK in magnetic field up to 7.5 T with low mechanical and electronic noise. The cryostat is not larger than conventional helium bath cryostats (23 and 13 liters of nitrogen and helium respectively), so that the setup does not require a large experimental hall and fits easily into a standard lab space. Mechanical vibrations with running dilution circulation were kept below 300 fm/$\sqrt{\mathrm{Hz}}$ by mechanically decoupling the STM from the cryostat and the pumping system. All electronic input lines were low-pass filtered, reducing the electronic temperature to below 100 mK, as deduced from the quasiparticle peaks of superconducting aluminium. The microscope is optically accessible in the parked position, making sample and tip exchange fast and user-friendly. For measurement the STM is lowered 60 mm down so that the sample ends in the middle of a wet superconducting magnetic coil.
Instrumentation and Detectors,Other Condensed Matter
What problem does this paper attempt to address?
The problem that this paper attempts to solve is the technical challenges faced in high - precision scanning tunneling microscope (STM) measurements at ultralow temperatures (millikelvin temperatures). Specifically, the author designed and constructed a compact ultra - high - vacuum scanning tunneling microscope (STM) that integrates a dilution refrigerator and can operate at extremely low temperatures (25 mK) and in strong magnetic fields (up to 7.5 T), while maintaining low mechanical and electronic noise. The following are the main technical challenges mentioned in the paper and their solutions: 1. **High - resolution measurement in a low - temperature environment**: - **Problem**: In order to study objects at the single - atom or molecular level, measurements need to be carried out at extremely low temperatures to prevent diffusion and improve energy resolution. - **Solution**: By integrating a dilution refrigerator, a substrate temperature of 25 mK is achieved, thereby improving the energy - spectrum resolution. 2. **Compact design**: - **Problem**: Traditional low - temperature STM systems are bulky, require large experimental spaces, and consume a large amount of liquid helium. - **Solution**: A compact cryostat was designed, whose volume is comparable to that of a traditional liquid - helium - bath cryostat and is suitable for standard laboratory spaces. Liquid - nitrogen and liquid - helium storage tanks are directly stacked in the ultra - high vacuum, reducing the overall size of the system. 3. **Low vibration and electronic noise**: - **Problem**: Mechanical vibration and electronic noise can interfere with the high - precision measurement of STM. - **Solution**: - By mechanically decoupling the STM from the cryostat and the pumping system, the mechanical vibration is reduced to below 300 fm/√Hz. - All electronic input lines are low - pass - filtered, reducing the electronic temperature to below 100 mK. 4. **Fast sample exchange**: - **Problem**: Sample exchange in traditional low - temperature STM systems is complex and time - consuming. - **Solution**: A fast sample - exchange mechanism was designed. The STM can replace samples and tips at optically visible positions, and the whole process is fast and user - friendly. 5. **Versatility and flexibility**: - **Problem**: Traditional low - temperature STM systems are single - function and difficult to adapt to various experimental requirements. - **Solution**: The system integrates multiple vacuum chambers and can perform in - situ sample preparation and surface analysis, while supporting various experimental conditions, such as different temperatures, magnetic fields, and sample types. 6. **Long life and low consumption**: - **Problem**: Traditional low - temperature systems consume a large amount of liquid helium and have a short running time. - **Solution**: The use efficiency of liquid nitrogen and liquid helium was optimized, reducing the liquid - helium consumption when the system is in steady - state operation and prolonging the running time. Through these innovative designs, the author successfully constructed a high - performance, compact low - temperature STM system, which is suitable for various scientific research, especially in solid - state physics research at the nanoscale.