Studies on adsorption-desorption of xenon on surface of BC-404 plastic scintillator based on soaking method

Xiang Yongchun,Fan Tieshuan,Zhang Chuanfei,Luo Fei,Wang Qian,Ze Rende,Xiang Qingpei
DOI: https://doi.org/10.1016/j.nima.2016.11.047
2017-03-01
Abstract:The phoswich coincidence detector is used to verify the CTBT treaty by measuring radioxenon and as such needs to possess high detection sensitivity. However, residual xenon adsorbed onto the surface of β detectors greatly influences subsequent measurements of weak samples. In this study, we investigate the adsorption-desorption behavior of xenon on BC-404 scintillator surfaces with different coating thicknesses using the soaking method. The results present the desorption behavior of xenon on a BC-404 surface for the first time. The calculated adsorption capacity for an uncoated surface is consistent with that from previous studies. However, due to factors such as limitations in coating technology, the effectiveness of coating on reducing the “memory effect” of the detector was poor. The proposed method is suitable for studying the adsorption-desorption behavior of gases on solid surfaces due to its simplicity and flexibility.
What problem does this paper attempt to address?