LIA-QMS method for the quantity analysis of the hydrogen isotopes retention in first-wall components of Globus-M2 tokamak

O.S. Medvedev,A.G. Razdobarin,E.V. Shubina (Smirnova),M.V. Grishaev,D.I. Elets,L.A. Snigirev,A.S. Kolomiytsev,A.V. Nikolaev
DOI: https://doi.org/10.1016/j.nme.2024.101829
IF: 3.037
2024-12-03
Nuclear Materials and Energy
Abstract:The technique based on quadrupole mass spectrometry of the material ejected by laser-induced ablation – LIA-QMS is proposed as a tool to measure H/D/T content in carbon co-deposits in the first wall and divertor regions of tokamak reactors. Tungsten tiles exposed in Globus-M2 tokamak were taken for validation experiments. Measurement accuracy was determined by comparison with the results obtained by laser-induced desorption – LID-QMS and conventional thermal desorption spectroscopy (TDS). Total amount of deuterium in hydrocarbon deposits was found to be 2.8 × 10 17 D/cm 2 , when measured by – LIA-QMS, 3.0 × 10 17 D/cm 2 for LID-QMS and 2.9 × 10 17 D/cm 2 for TDS. The main uncertainty of D surface concentration are caused by inhomogeneity of the deposit thickness over the sample area or by presence of hydrocarbon species in the released gas.
nuclear science & technology
What problem does this paper attempt to address?