Proof of principle for enhanced resolution multi-pass methods for the astral analyzer

Hamish Stewart,Dmitry Grinfeld,Bernd Hagedorn,Robert Ostermann,Alexander Makarov,Christian Hock
DOI: https://doi.org/10.1016/j.ijms.2024.117203
IF: 1.934
2024-01-30
International Journal of Mass Spectrometry
Abstract:Multi-pass zoom-mode methods have been described to extend the flight path and resolving power of multi-reflection time-of-flight mass spectrometers, by deflecting exiting ions back into repeat passes through the analyzer. Such methods have been adapted to the Astral analyzer, with repeated passes through a prism deflector found to compensate resolution-destroying aberrations. Changes to resolution were simulated and experimentally characterized, with more than a doubling of resolving power observed with only a moderate loss of transmission. Further methods are also proposed to eliminate the mass spectrum uncertainty caused by overlapping of signals from ions that have undergone unknown numbers of passes.
spectroscopy,physics, atomic, molecular & chemical