Crowd Control of Ions in the Astral Analyzer

Hamish Stewart,Dmitry Grinfeld,Johannes Petzoldt,Bernd Hagedorn,Michael Skoblin,Alexander Makarov,Christian Hock
DOI: https://doi.org/10.26434/chemrxiv-2023-p6zln
2023-12-28
Abstract:Space charge effects are the Achilles’ heel of all high-resolution ion optical devices. In time-of-flight mass analyzers these may manifest as reduction of resolving power, mass measurement shift, peak coalescence and/or transmission losses, whilst highly sensitive modern ion sources and injection devices ensure that such limits are easily exceeded. Space charge effects have been investigated, by experiment and simulation study, for the Astral multi-reflection analyzer, incorporating ion focusing via a pair of converging ion mirrors, and fed by a pulsed extraction ion trap. Major factors were identified as the resonant effect between ~103 ions of similar m/z in-flight, and the expansion of trapped packets of ~104-5 ions prior to extraction. Optimum operation and compensated ion mirror calibration strategies were then generated and described based on these findings.
Chemistry
What problem does this paper attempt to address?