A High Dynamic Range Ion Detector for the Astral™ Analyzer.
Hamish Stewart,Johannes Petzoldt,Toby Shanley,Dmitry Grinfeld,Eduard Denisov,Bernd Hagedorn,Ankit Dwivedi,Daniel Mourad,Robert Ostermann,Maximilian Ochmann,Philipp Cochems,Alexander Wagner,Wilko Balchun,Alexander Makarov,Christian Hock,Semyon Shofman,Ben-David Moti,Amit Weingarten,Sascha Kadyshevitch
DOI: https://doi.org/10.26434/chemrxiv-2024-49mzs
2024-05-28
Abstract:Reflectron-based time-of-flight analyzers rely on sub-nanosecond detector time response to achieve acceptable resolving power for low-mid mass, multiple ion peaks. With the adoption of multi-reflection analyzers, order of magnitude longer folded ion paths relax restrictions on detector response time, allowing implementation of new technologies that greatly improve dynamic range, detector lifetime, and ion detection efficiency. A detection system is presented, integrated into the Astral analyzer, that combines 10 keV post-acceleration and focal plane correction with a unique BxE focusing, optically coupled detector, pre-amplification and dual channel digitization. Calibration and peak handling methods are also described. The instrument demonstrated >1x104 dynamic range in a single shot, >100k resolving power, and a relative immunity to detector ageing.
Chemistry