Adaptive Proportional Derivative Control for Magnetic Bearing in Full Maglev Left Ventricular Assist Device

Kejia Zhang,Wenfei Tao,Chen Chen
DOI: https://doi.org/10.2478/msr-2024-0018
2024-08-01
Measurement Science Review
Abstract:In this paper, adaptive proportional derivative (APD) parameter control was proposed to solve the problem of high power consumption caused by the unclear mechanism of liquid disturbance during the lifting-up of magnetic bearing in left ventricular assist devices. A mathematical model was derived that describes how the rotor operates in liquid filling. The disturbance caused by the liquid in the lifting-up process was analyzed, and an adaptive control system was developed to improve dynamic performance and reduce power consumption. The experimental results show that APD control requires a shorter rise time without overshoot of rotor displacement compared to traditional fixed configurations. When using the APD controller, the peak current dropped by 8%. The duration in which the current is greater than 1A was reduced by 10.2 ms, and the average current also dropped by 34%.
Medicine,Engineering
What problem does this paper attempt to address?