130+ ps Coincident Time Resolution With 20mm Crystal Length Using 28 nm Xilinx FPGA

Fei Wang,Jiawen Zhou,Ziyi Weng,Chao Cai,Qingguo Xie
DOI: https://doi.org/10.1109/TRPMS.2024.3437178
2024-01-01
IEEE Transactions on Radiation and Plasma Medical Sciences
Abstract:The Coincidence Time Resolution (CTR) is of paramount importance in Positron Emission Tomography (PET) as it can directly determine the imaging resolution. In this paper, a 130+ ps CTR with 20 mm crystal length is achieved using AMD FPGA platform. Three steps are proposed to achieve a high CTR. First, a Low Noise Amplifier (LNA) is used on fast output signals that are used for time sampling. This can equivalently lower the configured threshold for leading edge detection and therefore further mitigate the time walk effect. Second, a new Time to Digital Converter (TDC) architecture that achieves less than 1 LSB Integral Non-Linearity (INL) and Differential Non-Linearity (DNL) without any calibration tricks are introduced. This TDC can yield salient INL performance, which can deliver consistent performance in time sampling and hence better CTR. Last but not least, a resource-efficient energy characterization method is proposed. This approach utilizes only one TDC chain to sample all trigger times for pulse reconstruction. This not only saves up to 75% chain resources but also minimizes sampling errors due to heterogeneity properties when involving multiple TDC chains. A prototype using 28 nm Kintex 7 FPGA is implemented and 130+ ps CTR is achieved.
What problem does this paper attempt to address?