High-precision Detection of Acetylene Using Photoacoustic Spectroscopy with a Three-Step Cylindrical Cell

Qiaoyun Wang,Lei Li,Shunyuan Xu,Ziheng Zhu,Jilong Wang,Xin Zou,Qiang Liu,Jibing Cheng,Richard Fu
DOI: https://doi.org/10.1109/jsen.2024.3398657
IF: 4.3
2024-01-01
IEEE Sensors Journal
Abstract:High-precision photoacoustic spectroscopy (PAS) with a three-step cylindrical photoacoustic cell (TSC-PAC) was developed for acetylene (C 2 H 2 ) detection. Finite element analysis was used to investigate the acoustic properties of the proposed TSC-PAC and conventional H-type cylindrical photoacoustic cell (HTC-PAC). Simulation results showed that acoustic pressure signal of the TSC-PAC was higher than that of HTC-PAC with the same dimension. Resonance frequency of the TSC-PAC was identified at 984 Hz. In experimental studies, a plane mirror was used to reflect the incident light and increase the optical length in the PAC. A silicon-cantilever-based fiber-optic sensor was positioned at the terminal point of the TSC-PAC, where the maximum photoacoustic (PA) signal was located. With C 2 H 2 concentration of 100 ppm, the 2 f signal of TSC-PAC was 1.43 times that of HTC-PAC. When the integration time was 100 s, the linear fitting (R 2 ) and minimum detection limit (MDL) of C 2 H 2 measured by TSC-PAC at the atmospheric pressure were 0.996 and 13.74 ppb, respectively. The normalized noise equivalent absorption (NNEA) coefficient was 8.18×10 -8 cm -1 ·W·Hz -1/2 . Both simulation and experimental results proved that the developed TSC-PAC enhanced the performance of the PAS and it had certain application potential in the PAS.
What problem does this paper attempt to address?