Abstract:Objective: To study biomechanical influence and relations between the elastic modulus and mechanical compatibility for prosthesis of different stiffness after replacement.Method: 21 fresh femoral specimen are used and divided into three groups:(1)normal group (2) comparative group of domestically made imitative AML CoCrMo alloy prosthesis and CRF\PSF composite material of the same geometry after the imitative femoral prosthesis replacement respectively.(3) experimemtal group,which makes use of electronic method to test the stress distribution,micro motion and torsional strength of femoral and prosthesis.Result:The stress and strain of femoral proximal tension that CRF\PSF group transmitted reaches 90%~95%,which is almost near to the natural physiological condition under stress; while CoCrMo group only gets to 60%~65%. For the micro motion of the prosthesis,CRF\PSF group is 30%~35% lower than CoCrMo group.The tortional strength of the former,however,is 30% higher than the latter. While for the stress concentration,CRF\PSF is 20%~30% lower than CoCrMo.Conclusion:CRF\PSF prosthesis of low elastic modulus can effectively reduce stress shielding and stress concentration. The early stability of prosthesis is much better than CoCrMo alloy.