Short carbon fiber reinforced polyetgeretherketone for hip prosthesis: Biocompatibility and biomechanics

Ke-jun Wang,Wei-chun Guo,Jin Tang,Ling Yu,Sheng-hao Zhao
DOI: https://doi.org/10.3969/j.issn.1673-8225.2011.34.020
2011-01-01
Abstract:Background: Polyetgeretherketone has good biocompatibility, but its modulus and mechanical strength characteristics drop badly upon glass-transition temperature, so as to fail to provide a steady condition for bone growth. Objective: To observe the biocompatibility and biomechanics of short carbon fiber reinforced polyetgeretherketone as hip prosthesis by animal experiment. Methods: The biocompatibility in vitro was measured by MTT, hemolysis test, acute toxicity experiment and pyrogen experiment; SCF/PEEK composite plates were implanted into the rabbits, and the formation of fibrous capsules and general tissues conditions around the plates were observed under the optical microscope; The stress tests were carried on the WE-SA universal testing machine. Results and Conclusion: The relative survival rate was > 75%, the hemolysis rate < 5%, indicating a good blood compatibility. There was no evident difference between the two groups. As the time goes on, the relation between the implanted-plates and capsules was more and more close, and inflammatory cells became gradually less.