Fluorimetric Determination of Arsenite and Arsenate in Water Using Fluorescein and Iodine

Bin Yuan,Jiuhui Qu,Qingzan Lin
DOI: https://doi.org/10.1080/03067310290024076
2002-01-01
International Journal of Environmental Analytical Chemistry
Abstract:A rapid, simple and sensitive fluorimetric method has been developed for the determination of inorganic arsenic. The method is based on the competitive reaction of arsenite and 2',7'-dichlorofluorescein (DCF) as fluoregenic reagent ( u ex = 510 nm, uem = 528 nm) at pH 6.5-7.5 with iodine (I2 ) to cause the increasing fluorescence intensity of the solution. Arsenate and other ions do not react with iodine. After reduction to the trivalent state by using L -cysteine, total arsenic is determined by spectrofluorimetry. Arsenate concentration can be calculated by difference. The calibration graph is linear over the range 4-180 ng/mL arsenite. The linear regression equation is j F = 7.82 C + 0.76 and the relative coefficient is 0.9991. The detection limit is 0.6 ng/mL arsenite. The recovery of this method for detecting arsenite is 96-105%. The method has been applied successfully to the determination of arsenite in tap and pool water.
What problem does this paper attempt to address?