The Energy Response of LaBr3(Ce), LaBr3(Ce,Sr) and NaI(Tl) Crystals for GECAM
Pei-Yi Feng,Xi-Lei Sun,Zheng-Hua An,Yong Deng,Cheng-Er Wang,Huang Jiang,Jun-Jie Li,Da-Li Zhang,Xin-Qiao Li,Shao-Lin Xiong,Chao Zheng,Ke Gong,Sheng Yang,Xiao-Jing Liu,Min Gao,Xiang-Yang Wen,Ya-Qing Liu,Yan-Bing Xu,Xiao-Yun Zhao,Jia-Cong Liu,Fan Zhang,Hong Lu
2023-12-28
Abstract:The GECAM series of satellites utilize LaBr3(Ce), LaBr3(Ce,Sr), and NaI(Tl) crystals as sensitive materials for gamma-ray detectors (GRDs). To investigate the non-linearity in the detection of low-energy gamma rays and address errors in the E-C relationship calibration, comprehensive tests and comparative studies of the non-linearity of these three crystals were conducted using Compton electrons, radioactive sources, and mono-energetic X-rays. The non-linearity test results for Compton electrons and X-rays displayed substantial differences, with all three crystals showing higher non-linearity for X-rays and gamma-rays than for Compton electrons. Despite LaBr3(Ce) and LaBr3(Ce,Sr) crystals having higher absolute light yields, they exhibited a noticeable non-linear decrease in light yield, especially at energies below 400 keV. The NaI(Tl) crystal demonstrated excess light output in the 6~200 keV range, reaching a maximum excess of 9.2% at 30 keV in X-ray testing and up to 15.5% at 14 keV during Compton electron testing, indicating a significant advantage in the detection of low-energy gamma rays. Furthermore, this paper explores the underlying causes of the observed non-linearity in these crystals. This study not only elucidates the detector responses of GECAM, but also marks the inaugural comprehensive investigation into the non-linearity of domestically produced lanthanum bromide and sodium iodide crystals.
Instrumentation and Detectors,Instrumentation and Methods for Astrophysics,High Energy Physics - Experiment,Nuclear Experiment